Topical: EU Commission wants to improve MDR
The time has come! From now on, the requirements of EU Regulation 2017/745 (MDR) apply.
In order to maintain the availability of medical devices on the EU market, the EU Commission plans to improvements to EU Regulation 2017/745 (MDR).
Our MDR Ticker (german) keeps you up to date. Or contact us by email. We will be happy to send you new information as soon as we have it.
With the help of the clinical evaluation, all manufacturers of medical devices (Class I to III) must demonstrate that the benefits, performance and safety of the devices are adequate. In other words, it must be ensured that the product does not compromise either the clinical condition or the safety of patients or third parties. Each medical device must therefore either still meet the essential requirements (ER) of the MDD or meet the essential safety and performance requirements (GSPR) of the MDR.
A clinical evaluation is composed of the following documents:
How the clinical evaluation should be structured and what content it must contain is described in numerous regulations and standards. These include, among others, Annex X of the Medical Device Directive (MDD); Articles 10 and 61, as well as Annex XIV of EU Regulation 2017/745 (MDR), Paragraph 19 of the Medical Devices Act (MPG) and MEDDEV 2.7.1.
In addition to the CEP and CER, the documentation must contain a number of other documents. The type and scope of this documentation depends on the class of the medical device (see figure). The risk management file, for example, should also be included here, since risk management is an essential prerequisite for clinical evaluation and vice versa.
Clinical data form the basis of any clinical evaluation. These data are used to evaluate whether the medical device is safe and performs well, and also allow the manufacturer to assess whether the risks encountered in connection with use are commensurate with the expected benefits of the device. The data can come from numerous sources, e. g.:
Notified bodies expect manufacturers to use multiple sources for clinical evaluation. However, the scope of the clinical evidence must be based on the characteristics of the product and its intended use. The scope for a standard Class I product from infection control, for example, may be less than for a Class III product. Manufacturers are therefore required to specify and justify the scope of the selected clinical data (clinical evidence). They must also demonstrate that the selected clinical data are sufficient to meet the essential safety and performance requirements. (Source: https://www.bundesgesundheits-ministerium.de/naki.html).
Neither the Medical Device Directive (MDD), nor the EU Regulation 2017/745 (MDR) allow the renouncing of a clinical evaluation. However, under certain conditions, clinical evaluation based on clinical data can be renounced. However, this usually only applies to absolutely non-critical products.
The MDR states in this regard in Article 61, paragraph 10:
Without prejudice to paragraph 4, where the demonstration of conformity with general safety and performance requirements based on clinical data is not deemed appropriate, adequate justification for any such exception shall be given based on the results of the manufacturer's risk management and on consideration of the specifics of the interaction between the device and the human body, the clinical performance intended and the claims of the manufacturer. In such a case, the manufacturer shall duly substantiate in the technical documentation referred to in Annex II why it considers a demonstration of conformity with general safety and performance requirements that is based on the results of non-clinical testing methods alone, including performance evaluation, bench testing and pre-clinical evaluation, to be adequate.
According to MDR, Article 61, Section 3, a clinical evaluation shall follow defined and methodologically sound procedure based on the following:
Since June 2016, the exact procedure has been described in detail in the clinical evaluation guideline published by the European Commission (MEDDEV 2.7/1 rev. 4). A new feature is that the clinical evaluation plan, which defines the objectives and structure of the clinical evaluation, must in future also include a clinical development plan.
The clinical development plan, for example, describes in detail the clinical planning envisaged by the manufacturer in order to be able to demonstrate the clinical safety, the lowest possible burden and the effective benefit of the medical device to be evaluated via its own clinical data. Clinical planning ranges from exploratory to pivotal studies to post-marketing clinical follow-up (PMS), specifying milestones and describing possible acceptance criteria.
In general, according to the MDR, a clinical evaluation and associated documentation must be updated throughout the lifecycle of the device based on clinical data. For this purpose, the manufacturer needs a plan for the Post Market clinical follow-up in accordance with Annex XIV Part B of the MDR and Article 84 (see PMCF studies).
Briefly described, a clinical evaluation proceeds as follows:
According to the MDR, the concept of similarity to other products can still be applied to products for which clinical data are already available, but only in a limited number of situations. The new rules are also stricter (MDR, Article 61, 3 MDR, and Annex XIV, paragraph 3). They state that:
A clinical evaluation may be based on clinical data relating to a device for which equivalence to the device in question can be demonstrated. The following technical, biological and clinical characteristics shall be taken into consideration for the demonstration of equivalence:
The characteristics listed in the first paragraph shall be similar to the extent that there would be no clinically significant difference in the safety and clinical performance of the device. Considerations of equivalence shall be based on proper scientific justification. It shall be clearly demonstrated that manufacturers have sufficient levels of access to the data relating to devices with which they are claiming equivalence in order to justify their claims of equivalence.
MDCG 2020-5 "Clinical Evaluation - Equivalence" can be used as a guideline by manufacturers and notified bodies to perform equivalence in accordance with the standard. In general, the requirements listed in the MDR (biological, technical, clinical) should be read very carefully, as the requirements can only be met in part if the same aspects are fulfilled or even if they are only similar.
In cases where equivalence cannot be demonstrated under the MDR, the data from similar devices may be useful for a variety of other purposes, for example (MDCG 2020-5):
Please note that the MDCG guidance documents do not have a binding character for authorities or notified bodies. However, current practice shows that the suggestions of these documents are followed in the vast majority of cases.
As the MDR severely restricts equivalence assessment, this means that you, as a manufacturer, will need to generate and maintain more of your own clinical data for your products in order to be able to show clinical evidence with the essential safety and performance requirements as part of the clinical evaluation. You will need to account for this accordingly in the clinical development plan.
In addition, due to the limitations of equivalence consideration and the stringent requirements for the quality of clinical data, clinical trials for the initial marketing of a medical device and PMCF studies for recertification will be increasingly required. For Class III devices and implantable devices, these are even mandatory to perform, with a few exceptions (see below).
Therefore, as a manufacturer, you should critically review your own clinical data for the devices in this regard. If you identify gaps or find that your products are classified differently than before, or if you have to clinically evaluate your products for the first time, you should use the time until the end of May 2021 to actively generate clinical data on your own products via clinical trials or PMCF studies and to perform clinical evaluations.
The publications of the EU Medical Device Coordination Group (MDCG) provide some relief for manufacturers of legacy medical devices. In particular, MDCG 2020-6 and MDCG 2021-25 specify the applicable sources for clinical data for such "legacy devices". The requirements of the MDR for the PMS and, if necessary, for the PMCF must nevertheless be met in full.
For existing devices, the so-called legacy devices, the MDCG has published the document MDCG 2020-6: "Regulation (EU) 2017/745: Clinical evidence needed for medical devices previously CE marked under Directives 93/42/EEC or 90/385/EEC". It is intended to serve as a guide for clinical data sufficient to provide evidence of compliance with the relevant GSPRs under Article 61(1) of the MDR for legacy devices.
Again, it should be noted that these guides are not binding on authorities or notified bodies.
According to MDCG 2021-25, legacy devices should be understood as devices, which, in accordance with Article 120(3) of the MDR, are placed on the market after the MDR’s date of application (DoA) and until 26 May 2024 if certain conditions are fulfilled. Those devices can be:
‘Old’ devices are those devices that were placed on the market before 26 May 2021 in accordance with the AIMDD or the MDD or in accordance with the applicable rules before the Directives had entered into force (MDCG 2021-25).
The terminology “Well-established technology” is used in Article 52(5) and Article 61(8) of the MDR, but is not defined in these articles. The term is not restricted to the devices listed in Article 61(6b); Article 61(8) explicitly states that this includes devices similar to the exempted devices listed in Article 61(6b), which might be added to that list in future. The common features of the devices which are wellestablished technologies are that they all have:
When assessing the conformity of legacy devices under the MDR, it is important to verify whether PMCF studies considered necessary under the MDD/AIMDD (and where applicable, during the transition period, under the MDR), have been appropriately conducted, and results are taken fully into account for in the clinical evaluation for the conformity assessment under MDR (MDCG 2020-6).
Manufacturers are required to document a clinical evaluation plan to meet the requirements of MDR Annex XIV Section 1a.
A modified Clinical Evaluation Plan for legacy devices should include at least:
For the purpose of legacy devices, pre-market sources of clinical data may include (MDCG 2020-6):
It should be noted that MDR Article 2(48) provides a narrower definition of what constitutes clinical data sources as compared to the Directives which allow unpublished reports on other clinical experience to contribute to the clinical evaluation. Such data sources may provide informative context for the clinical evaluation of legacy devices.
Post-market sources of clinical data refer to data collected following the initial CE marking under the Directives (or prior to introduction of a new indication or design variant). This may include (MDCG 2020-6):
For well-established technologies the clinical evaluation can be based on data coming from similar devices, under the conditions detailed in paragraph 6.5 (e). With respect to legacy devices, when clinical data from equivalent devices is used, equivalence must be demonstrated according to the requirements of the MDR.
MDCG 2020-6 also proposes a hierarchy of clinical evidence to confirm compliance with the relevant GSPRs under the MDR, see Appendix II of the guidance document.
According to MDR Artikel 61, 4 in the case of implantable devices and class III devices, clinical investigations shall be performed, except if:
In this case, the notified body shall check that the post-market clinical follow-up plan is appropriate and includes post-market studies to demonstrate safety and performance of the device (MDR, Article 61, paragraph 4).
Article 61, paragraphs 5 and 6 state, that a manufacturer of a device demonstrated to be equivalent to an already marketed device not manufactured by him, may also rely on paragraph 4 in order not to perform a clinical investigation provided that the following conditions are fulfilled in addition to what is required in that paragraph:
and the manufacturer of the second device provides clear evidence thereof to the notified body.
It is further valid that the requirement to perform clinical investigations pursuant to paragraph 4 shall not apply to implantable devices and class III devices,
a) which have been lawfully placed on the market or put into service in accordance with Directive 90/385/EEC or Directive 93/42/EEC and for which the clinical evaluation:
b) that are sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips or connectors for which the clinical evaluation is based on sufficient clinical data and is in compliance with the relevant product-specific CS, where such a CS is available.
In general, it is now only possible to dispense with a clinical evaluation under very specific conditions. Article 61, paragraph 10 of the MDR states in this regard:
Without prejudice to paragraph 4, where the demonstration of conformity with general safety and performance requirements based on clinical data is not deemed appropriate, adequate justification for any such exception shall be given based on the results of the manufacturer's risk management and on consideration of the specifics of the interaction between the device and the human body, the clinical performance intended and the claims of the manufacturer. In such a case, the manufacturer shall duly substantiate in the technical documentation referred to in Annex II why it considers a demonstration of conformity with general safety and performance requirements that is based on the results of non-clinical testing methods alone, including performance evaluation, bench testing and pre-clinical evaluation, to be adequate.
In principle, you as a manufacturer may now only waive a clinical evaluation for completely uncritical medical devices.
Sources:
Regulatory affairs for clinical evaluation
Clinical evaluation is a systematic and ongoing process that plays a role both in the product development phase and after the product has entered the market, right up to the end of the product life cycle.
In the context of clinical evaluation, we can therefore support you with the following services, so that your product receives approval quickly and your documentation is always up to date, even after market entry:
With our help you can:
If we have aroused your interest, please contact us. Call us, write to us or fill out our contact form. We look forward to hearing from you.
Regulatory affairs for clinical evaluation
We support you as a manufacturer in every phase of the clinical evaluation of your medical device from setting up the clinical evaluation plan, to performing II a, IIb, b, III on.
This includes the following services:
For this purpose, we use our own forms created according to MedDev 2.7/1 Rev. 4, which we are also happy to provide to you. These were created according to the guidelines of DIN EN ISO 9001, which means that we regularly check and update them.
Generally, we offer our services in German as well as in English.
Contact us and we will be happy to create a customized offer for you.
Clinical evaluation is a methodologically sound, ongoing process for collecting, assessing and analyzing clinical data on a medical device. The purpose is to demonstrate that the benefits, performance and safety of the medical devices are adequate.
A clinical evaluation consists of, among other things, the clinical evaluation plan including the clinical development plan (CEP) and the clinical evaluation report (CER). There should also be a literature search plan and report either as a separate document or as part of the CEP and CER.
The clinical evaluation should be conducted by an appropriately qualified person or team. The person or team should have, among other things, experience in research methodology and regulatory requirements, have, for example, knowledge of product technology and expertise in the relevant medical field, and have appropriate training (studies, work experience) (See MDR, Article 15 or MEDDEV 2.71 / rev.4, Section 6.4)